Geoduck homogenization and RNA extraction
20160824-20160829 Homogenization of juvenile geoduck samples for use in nucleic acid extractions
Tissue Homogenization for Nucleic Acid Analysis Protocol
Sample list 20160824
- EPI-276
- EPI-282
- EPI-283
- EPI-289
- EPI-290
- EPI-298
- EPI-299
- EPI-302
- EPI-303
Sample list 20160825
- EPI-275
Homogenized samples on liquid nitrogen and split in half for RNA and DNA extractions Placed samples immediately on dry ice and moved them directly to -80° storage
RNA Extractions
- Added 1.0ml of RNAzol to each tube and homogenized with teflon pestle
- moved 0.5ml of homogenate to new tube
- Added 0.4ml of nuclease free water to each new tube
- Added 0.5ml of RNAzol to the initial homogenate-RNAzol mix and replaced at -80°C
- Proceeded with RNA Extraction Protocol
RNA Quantification 20160826
- Diluted samples 1:100 in nucelase-free water
- Used 10µl of sample and 190µl of Qubit Mix
- Ran RNA Quantification Protocol
RNA Concentrations
Sample.ID | Conc.(ng/µl) |
EPI-275 | 582 |
EPI-276 | 1180 |
EPI-282 | 1160 |
EPI-283 | 1160 |
EPI-289 | 790 |
EPI-290 | 1060 |
EPI-298 | 1020 |
EPI-299 | 762 |
EPI-302 | 888 |
EPI-303 | 706 |
RNA Quality 20160826
- Diluted above 1:100 dilution of samples again 1:10 and ran on bioanalyzer (plus test sample EPI 291)
- Used samples from above 1:100 dilute and ran on bioanalyzer (plus test sample EPI 291)
- Results were very noisy and degraded, ladder did not display correctly despite the fact it was new
- Appears to be a problem with chip, kit, ladder, or setup
RNA Quality 20160829
- Issue with 8-26 Bioanalyzer was identified as wrong clip position on loading syringe. Moved clip to top position.
- Diluted samples extracted on 20160824 - 20160825 1:100 and ran on bioanalyzer along with a positive control from larval geoduck RNA extracted by SJW (#35)
- BioAnalyzer 20160829 Chip 1
Conclusions
- Only a single large peak is expected for geoduck RNA, as the in the rRNA the 28S contains a hidden break that when the sample is heat denatured will break the 28S in half and these fragments will co-migrate with the 18S
- Many peaks seen prior to the tall single peak, some potential for genomic DNA to be present after the large rRNA peak
- The identical patterns of the smaller peaks prior to the large rRNA peak suggests the small peaks are not due to degradation. It is possible they are RNAs from other “symbiotic” or prey organisms as the geoducks are fed diatoms and dinoflagellates. The positive control would not have these small peaks as the prey is in too low of concentration at the larval stage the RNA was obtained from.
- To check for degradation and to minimize genomic carry over, RNA extraction should be repeated with a higher RNAzol to tissue ratio to minimize genomic DNA carryover and to prevent the capacity of the spin columns from being overwhelmed.
Written on August 30, 2016